tiamo 2.3 Patch 1

■■■■■■■■

775

Titrino
Range

-170.0 ... 500.0 °C

Default value

25.0 °C

Time interval measuring point

Time interval for entering a measuring point in the measuring point list.

Titrando, 855
Range

0.1 ... 999999 s

Default value

2.0 s

Titrino
Range

0.08 ... 16200 s

Default value

2.00 s

5.6.3.4.6.6

SET Upol - Stop conditions

Tab: Method

SET Upol Properties... Stop conditions

Command name

Name of the command.

Input

25 characters

Conditions for stopping the titration. If more than one stop condition is in
effect at the same time, then the criterion which is fulfilled first will stop
the titration.

Stop volume

Stops when the given volume has been added after the start of the titra-
tion (including start conditions). The stop volume should be adapted to
suit the sample weight or the titration vessel size.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

100,000 mL

Selection

off

Titrino
Range

0.000 ... 9999.99 mL

Default value

100.00 mL

Selection

off

Stop time

Stops when the preset time (including start conditions) has elapsed since
the start of the titration.