5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1020

■■■■■■■■

tiamo 2.3 Patch 1

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

856
Range

0.0000 ... 2000.00 mS/cm

Selection

off

Default value

off

off
No stop.

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Duration of measurement

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.
Max. number of entries?

712
Range

0.4 ... 999999 s

Default value

2.0 s

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0.0000 ... 2000.00 mS/cm

Selection

off

Default value

off

off
No stop.