5 Method

tiamo 2.3 Patch 1

■■■■■■■■

873

Titration rate

Titration rate

Three predefined rates, for which the parameters for Control are not dis-
played, can selected for the titration rate. The User setting must be selec-
ted in order to be able to edit these parameters.

Selection

50 µL/min | 100 µL/min | 500 µL/min | User

Default value

100 µL/min

Control

Note

Will be displayed only if the option User is selected in the selection list
Titration rate.

Dynamics pH

The control range defines the measured value range before the given con-
trol point. It has a decisive influence on the titration rate and thus also on
the accuracy. Within the control range, dosing is controlled by the Min.
rate
. The closer the control point, the slower the dosing until the Min.
rate
has been reached. The larger the control range, the slower the titra-
tion. Outside the control range, dosing is carried out continuously at
Max. rate.

Titrando, 855
Range

0.001 ... 20.000

Default value

1.000

Selection

off

Titrino
Range

0.01 ... 20.00

Default value

1.00

Selection

off

Max. rate

Rate at which dosing is carried out outside of the Control range. The
maximum dosing rate depends on the cylinder volume of the exchange
unit or dosing unit used. The following rule of thumb should be used as a
starting point for parameterization: Max. rate in mL/min = 0.005 •
Expected rate of reaction in µL/min.