3 dos u - dosing parameters, Dosing parameters – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1203

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1187

5.6.6.3.5.3

DOS U - Dosing parameters

Tab: Method

DOS U Properties... Dosing parameters

Command name

Name of the command.

Input

25 characters

The volume to be added, the dosing rate and the dosing time are defined
on this tab.

Dosing criterion

Selects the two parameters that can be preset for dosing. The third para-
meter is calculated automatically from the formula Volume = dosing
time * dosing rate
and is not shown (see Chapter 5.6.6.3.2, page
1160)
.

Selection

Volume/dosing rate | Volume/dosing time |
Dosing rate/dosing time

Default value

Volume/dosing rate

Volume

Fixed volume that is to be dosed. This parameter is only displayed for
Dosing criterion = Volume/Dosing rate or Volume/Dosing time.

Range

0.00000 ... 99999.9 mL

Default value

10.0000 mL

Dosing rate

Rate at which dosing should take place. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing device
used. If the entered dosing rate is too high for the selected dosing device,
it will automatically be reduced during dosing to the largest possible
value. This parameter is only displayed for Dosing criterion = Volume/
Dosing rate
or Dosing rate/Dosing time.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Default value

maximum mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Advertising