5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

538

■■■■■■■■

tiamo 2.3 Patch 1

tionally meet the parameters defined for each window will be recognized.
Only one endpoint will be recognized per window.

Lower limit

Lower limit for the volume window.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Upper limit

Upper limit for the volume window.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Titrando, 855
Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.