Note

The immediate stop of all ongoing determinations at all active workpla-
ces with the emergency stop button remains possible, even after an
automatic logout.

Password policies

Note

If the option Password monitoring by Windows is switched on,
then the parameters for password protection are inactive. Compliance
with the regulations in accordance with 21 CFR Part 11 must then be
ensured by the administrator in Windows.

Enforce unique password

If this check box is activated then this ensures that a user can use a given
password only once.

Selection

on | off

Default value

off
FDA default: on (can no longer be edited)

Enforce use of special characters

If this check box is activated then this ensures that the password must
contain at least one special character (@, #, ~ etc.).

Selection

on | off

Default value

off
FDA default: off (editable)

Minimum password length

If this check box is activated then this ensures that the password must
contain the indicated number of characters.

Selection

on | off

Default value

off
FDA default: on (editable)

Range

1 ... 10 characters

Default value

6 characters
FDA default: 6 characters (editable)