4 Database

tiamo 2.3 Patch 1

■■■■■■■■

383

Time [s]

Time in s until limit violation.

Limit

Selection

lower | upper | ok

lower
Lower limit value exceeded.
upper
Upper limit value exceeded.
ok
Measured value within the limits again.

[pH]/[U]

Limit for the measured value.

Meas.v. [pH]/Meas.v. [U]

Measured value at limit violation time.

Volume [mL]

Dosed volume at limit violation time.

Temp. [°C]

Temperature at limit violation time.

Action

Selection

Cancel determination | Cancel command
(Titrando only)
| Wait for [Continue] | Wait for
limit ok
| none

Cancel determination
The running STAT command has been canceled, then the determina-
tion has been finished.
Cancel command (Titrando only)
The running STAT command has been canceled, then the next com-
mand has been carried out.
Wait for [Continue]
Reagent dosing in the running STAT command has been interrupted
until it has been continued by a manual command.
Wait for limit ok
Reagent dosing in the running STAT command has been interrupted
until the monitored measured value has been within the limits again
(including hysteresis). Afterwards reagent dosing has been resumed
automatically.