4 set u - control parameters – Metrohm tiamo 2.3 Patch 1 User Manual

Page 742

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

726

■■■■■■■■

tiamo 2.3 Patch 1

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Pause 2

Pause 2

Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the addition of the start volume. The pause follows at
the end of all the start conditions.

Range

0 ... 999999 s

Default value

0 s

5.6.3.4.4.4

SET U - Control parameters

Tab: Method

SET U Properties... Control parameters

Command name

Name of the command.

Input

25 characters

The control parameters can be set for each endpoint separately.

Endpoint 1

EP1 with

Measured value for the first endpoint.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Selection

off

Default value

off

Titrino
Range

-2000 ... 2000

Selection

off

Default value

off

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for Control are not dis-

Advertising