Metrohm tiamo 2.3 Patch 1 User Manual

Page 1116

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1100

■■■■■■■■

tiamo 2.3 Patch 1

Min. waiting time

The measured value is not accepted until the minimum waiting time has
elapsed, even if the signal drift has already been reached. The drift contin-
ues to be checked while the waiting time is elapsing.

856
Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If the signal drift has not yet been reached, then the measured value will
be accepted when the maximum waiting time has elapsed. If the waiting
time has not been newly entered then a waiting time that is suitable for
the drift will be calculated automatically according to the following equa-
tion:

856
Range

0 ... 999999 s

Default value

52 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.

856
Range

0.1 ... 999999.0 s

Default value

2.0 s

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Measuring time

Maximum period of time for measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measuring point in the measuring point list.
Max. number of entries?

712
Range

0.4 ... 999999 s

Default value

2.0 s

Advertising