Metrohm tiamo 2.3 Patch 1 User Manual

Page 1148

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1132

■■■■■■■■

tiamo 2.3 Patch 1

Measuring interval

Time interval for entering a measured value in the measuring point list.

Avantes Spectrometer
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

089 Photometer
Range

0.25 ... 999999.00 s (Increment: 0.25)

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off

Measurement without drift control
If the option Measurement without drift control has been selected,
then the following parameters will be displayed:

Duration of measurement

Duration of the adsorption measurement.

Range

0 ... 999999 s

Default value

120 s

Measuring interval

Time interval for entering a measured value in the measuring point list.

Avantes Spectrometer
Range

0.1 ... 999999.0 s (Increment: 0.1)

Default value

2.0 s

089 Photometer
Range

0.25 ... 999999.00 s (Increment: 0.25)

Default value

2.0 s

Stop measured value

Stop when the entered measured value has been reached since the start
of the measurement.

Range

0 ... 4000 mAU

Selection

off

Default value

off

Advertising