Metrohm tiamo 2.3 Patch 1 User Manual

Page 31

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

1 Introduction

tiamo 2.3 Patch 1

■■■■■■■■

15

With the command MEAS Cond, certain variables with only 2 decimal
points were identified in the database in the subwindow Results on
the tab Raw data, whereas the measured values with 4 decimal pla-
ces were displayed in the measuring point list.

The backup did not function when recovery was switched off. As a
new feature, this will be checked and reactivated each time a database
is opened in the event that it was switched off.

The modification of a method variable, for example, led to the mes-
sage 015-149 Missing data and to a stop when a pH calibration was
being recalculated.

It was not possible to restore databases from tiamo 1.x.

With the command MEAS Upol the command variable EME was out-
put with only one decimal point in the raw data.

Inconsistent data sometimes arose in isolated incidents. Possible symp-
toms of this were a database which began to slow down and a mas-
sive memory consumption.

Program part Method

The non-existent variable ERC was removed from the MET commands
(it was always invalid) and replaced by DME (measured value differen-
tial).

It sometimes happened with thermometric titration that the first deriv-
ative ran through a steep rise at the beginning, which meant that the
stop criterion Stop slope was already fulfilled. In order to avoid this,
the field Stop slope active after was introduced on the tab Stop
conditions
in connection with the command TET.

The field Flow rate was absent with the method parameter report for
the command FLOW.

If the device assignment was modified in connection with the com-
mand LQH, after the port had been redefined on the tab Parameters,
then the port changed back to the default value.

The selection of the thermoprobes was listed twice under Sensor in
connection with the command TET.

Program part Configuration

It sometimes happened that firmware updates could not be success-
fully completed (the update process came to a halt).

If the axial distance was redefined and tiamo shut down and restarted
with sample changers, then the axial distance was reset to the previous
value.

If Password monitoring by Windows was set in the security set-
tings and if the check box Maximum number of login attempts
was activated, then the number of incorrect login attempts was not
updated at the time of Windows login.

The measuring input was not implemented correctly in the configura-
tion with non-intelligent sensors.