5 Method

tiamo 2.3 Patch 1

■■■■■■■■

675

Designation

Description

.EP{x}.TIM

Time in s until the endpoint × (1…9) is reached

.EP{x}.VOL

Volume for the endpoint × (1…9) in mL

.ETE

End temperature (temperature following processing
of the command) in °C

.EVT

End volume (total dosed volume at the end of the
command) in mL

.FIN

Command status; 1 = Command has ended at
least once
, 0, invalid (variable not available) =
Command has never ended

.FP{x}.DME

Measured value differential for the fixed endpoint ×
(1…9) in the unit of the measured value

.FP{x}.MEA

Measured value for the fixed endpoint × (1…9) in
mS/cm

.FP{x}.TEM

Temperature for the fixed endpoint × (1…9) in °C

.FP{x}.TIM

Time in s until the fixed endpoint × (1…9) is
reached

.FP{x}.VOL

Volume for the fixed endpoint × (1…9) in mL

.IME

Initial measured value (measured value before start
conditions are processed) in the unit of the meas-
ured value

.ITE

Initial temperature (temperature before start condi-
tions are processed) in °C

.LP.CAx

Calculated value × (1…3) for the last measuring
point on the measuring point list

.LP.DME

Measured value differential for the last measuring
point on the measuring point list in the unit of the
measured value

.LP.EXx

External value × (1…3) for the last measuring point
of the measuring point list

.LP.MEA

Measured value for the last measuring point of the
measuring point list in the unit of the measured
value

.LP.TEM

Temperature for the last measuring point of the
measuring point list in °C