5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1229

Note

It is advisable to define an initial temperature if you work always at the
same temperature.

Heating time

The target temperature has to be reached within this time. The heating
rate can be modified with the heating time in order to program tempera-
ture ramps or entire temperature profiles (with several HEATER com-
mands). Off causes heating at the maximum possible heating rate (see
device manual).

This value is live editable for the 874

Range

1 ... 999 min

Selection

off

Default value

off

Max. waiting time

If the Target temperature is not reached within the defined Heating
time
then this waiting time will start. This parameter is only visible if a
Heating time has been defined.

This value is live editable for the 874.

Range

1 ... 999 min

Selection

off

Default value

off

Action if timeout expires

on | off (Default value: on)

If this option is activated, then an action can be defined which is to be
carried out if the target temperature has not yet been reached after expiry
of the maximum waiting time (is displayed only when Timeoutoff).

Selection

Display message | Cancel determination | Can-
cel determination and series

Default value

Display message

Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it.
Cancel determination
The determination is canceled and the next determination of the series
is started.