5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

560

■■■■■■■■

tiamo 2.3 Patch 1

Slope

Minimum difference between the slope before and after the break-point.
The smaller the difference, the more break-points will be found.

Range

0.0 ... 10.0

Default value

0.9

Smoothing factor

The higher the smoothing factor, the fewer endpoints will be found.

Range

2 ... 20

Default value

5

Window

A range (window) can be defined on the measured value axis, on the vol-
ume axis or on the time axis. The break-point evaluation will only be car-
ried out in the defined window. Only the first break-point in the defined
window will be recognized.

Selection

Measured value | Volume | Time | off

Default value

off

Lower limit

Value for the lower limit of the window.

Window = Measured value
Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Window = Volume
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Window = Time
Range

0 ... 999999 s

Default value

0 s

Upper limit

Value for the upper limit of the window.

Window = Measured value
Range

-200.0 ... 200.0 µA

Default value

200.0 µA

Window = Volume
Range

0.00000 ... 9999.99 mL

Default value

9999.99 mL