5 calibration commands, 1 calibration commands - overview, Calibration commands 9 – Metrohm tiamo 2.3 Patch 1 User Manual

Page 1105

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1089

5.6.5

Calibration commands

5.6.5.1

Calibration commands - Overview

Menu item: Method

Insert New command... Calibrate

Commands for calibration of pH- and ion-selective sensors.

General
The commands CAL LOOP and CAL MEAS are available for calibrating
electrodes. The number and type of buffers or standards, respectively, is
defined in the loop command CAL LOOP, which is to be measured sub-
sequently with the command CAL MEAS. After each looping pass, the
validity of the measurement is checked, and at the end the calibration
data for the electrode is calculated.

The basic structure of a calibration method looks as follows:

Advertising