5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

796

■■■■■■■■

tiamo 2.3 Patch 1

Stop drift

Titration is stopped when the endpoint and the stop drift have been
reached. This parameter is only shown for Stop criterion = drift.

Range

1 ... 999 µL/min

Default value

20 µL/min

Relative stop drift

The titration is stopped when the endpoint and the stop drift have been
reached. The stop drift is the sum of the drift at the titration start and the
relative stop drift defined here. This parameter is only shown for Stop cri-
terion
= rel. drift.

Titrando
Range

1 ... 999 µL/min

Default value

10 µL/min

Delay time

If the endpoint has been reached, the delay time defined here is awaited
after the last dosing and then the titration is stopped. This parameter is
only shown for Stop criterion = time. The selection inf. stands for infin-
ite.

Titrando, 855
Range

0 ... 999 s

Default value

10 s

Titrino
Range

0 ... 999 s

Default value

10 s

Selection

inf.

Stop time

The titration is stopped if the absolute stop time defined here has elapsed
since the start of the titration. This parameter is only shown for Stop cri-
terion
= time and Delay time = inf.

Titrino
Range

1 ... 999999 s