5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

670

■■■■■■■■

tiamo 2.3 Patch 1

[Properties]

Open the dialog window External measured value # in which the
parameters for the external measured value can be edited (see Chapter
5.6.3.11.2, page 931)
.

[Delete]

Delete the external measured value selected in the table.

5.6.3.3.5.9

MET Upol - Measured value window

Dialog window: Method

MET Upol Properties... Potentiometric evalua-

tion

Evaluation with measured value window [New]/[Properties] Meas-

ured value window#

Measured value windows are regions on the measured value axis for
which different parameters for potentiometric evaluation can be defined.
Only those endpoints which lie within these windows and additionally
meet the parameters defined for each window will be recognized. Only
one endpoint will be recognized per window.

Lower limit

Lower limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

-200.0 µA

Upper limit

Upper limit for the measured value window.

Range

-200.0 ... 200.0 µA

Default value

200.0 µA

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Range

0.1 ... 99.9 µA

Default value

2.0 µA

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.