5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1199

Compensate
As the dosing units are exchangeable, the coupling of the Dosino pis-
ton rod (spindle) has a low mechanical tolerance that can be noticed
when the piston changes its direction of movement. This tolerance can
be compensated for with the Compensate function. A short piston
movement is first made in the same direction as the previous move-
ment, which is then followed by a piston movement in the reverse
direction.

Port

The port where each function is to be carried out needs to be defined for
each liquid handling command.

Selection

1 | 2 | 3 | 4

1
Default setting for Dosing and Aspirate
2
Default setting for Fill, Exchange position and Change port
4
Default setting for Eject to stop, Eject to end volume and Compensate

Volume

Enter the volume to be moved. This parameter is available only for the
Functions Dosing or Aspirate.

Dosing

Range

0.00000 ... 99999.9 mL

Default value

10.0000 mL