5 Method

tiamo 2.3 Patch 1

■■■■■■■■

905

Lower hysteresis

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok
is triggered.

Titrando, 855
Range

0.0 ... 2000.0 mV

Default value

0.2 mV

Upper limit

Upper limit of the measured value. If the measured value exceeds this
limit, then the event Measured value upper limit exceeded is trig-
gered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

Titrino
Range

-2000 ... 2000

Default value

2000

Upper hysteresis

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0.0 ... 2000.0 mV

Default value

0.2 mV

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Cancel determination | Cancel command
(Titrando, 855)
| Wait for [Continue] | Wait for
limit ok
| none

Default value

none

Cancel determination
The running STAT command will be canceled, then the exit track (if
present) will be started and the determination will be finished.
Cancel command (Titrando, 855)
The running STAT command will be canceled, then the next com-
mand will be carried out.