5 Method

tiamo 2.3 Patch 1

■■■■■■■■

835

The diagram shows KF titration curves at different polarization currents. It
is clear to see that the position of the endpoint varies with the polarization
current. The curves have different slopes, i.e. dynamics must also be adap-
ted. Polarization currents smaller than 10 µA are not suitable for this appli-
cation. The following table gives an idea of the optimal control parame-
ters for various polarization currents.

Ipol

10 µA

20 µA

30 µA

EP

50 mV

100 mV

150 mV

Control range

70 mV

100 mV

120 mV

Max. rate

Maximum rate with which iodine is generated outside the Control
range
.

Range

1.5 ... 2240.0 µg/min

Default value

1000.0 (slow), maximal (optimal), maximal
(fast) µg/min

Selection

maximum

Min. rate

This parameter defines the minimum rate with which iodine is generated
inside the Control range. The smaller the rate the slower the titration
and the higher the accuracy.