6.8 Subwindow Colorimetric sensors

■■■■■■■■■■■■■■■■■■■■■■

1424

■■■■■■■■

tiamo 2.3 Patch 1

Calibration interval

Validity of the calibration in days. If a value is entered here then the date
for the next calibration will be adapted automatically. The days are
monitored, e.g. the calibration becomes invalid when the date changes
and not at the time time of day that calibration took place.

Only editable for Calibration data monitoring = on, otherwise inactive.

Range

0 ... 999 Days

Default value

999 Days

Next calibration

Date on which the next calibration is to be carried out. The date can be

selected by pressing

in the dialog window Select date. After the date

has been entered the value for Calibration interval will be automatically
adapted.

Only editable for Calibration data monitoring = on, otherwise inactive.

Default value Calibration date+ 999 days

Message

Message by e-mail

on | off (Default value: off)

The message is additionally sent to the address defined under [E-mail...]
if this check box is activated.

Editable only if Calibration data monitoring = on, otherwise inactive.

[E-mail...]

[E-mail...] opens the Send e-mail window (see Chapter 2.6.1, page 95).

Active only if Message by e-mail = on.

Acoustic signal

on | off (Default value: off)

If this check box is activated, an acoustic signal will be emitted in addition
to the message defined above.

Editable only if Calibration data monitoring = on, otherwise inactive.

Action
If during calibration data monitoring it is found that its validity has
expired, then one of the following actions will be triggered automatically
at start test:

Editable only if Calibration data monitoring = on, otherwise inactive.