Metrohm tiamo 2.3 Patch 1 User Manual

Page 855

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 Patch 1

■■■■■■■■

839

off
No drift correction takes place.

Drift value

Drift value for drift correction. This parameter is only shown for Drift cor-
rection
= manual.

Range

0.0 ... 99.9 µg/min

Default value

0.0 µg/min

Stop time

Determination stops when the specified time has elapsed since the start of
the conditioning.

Range

1 ... 999999 s

Selection

off

Default value

off

Stabilizing time

Waiting period until the start of the titration during which the condition
requirement must be continuously fulfilled.

Range

0 ... 999 s

Default value

0 s

Automatic start after sample addition time

on | off (Default value: on)

If this option is activated, then the titration will be started automatically
after the sample addition time.

Sample addition time

Waiting time between stop of conditioning and start of titration during
which the sample can be added.

Range

0 ... 99 s

Default value

0 s

Manual start after sample addition with [Continue]

on | off (Default value: off)

If this option is activated, then the titration will be started manually after
the sample addition by pressing the [Continue] button in the message
window.