Metrohm tiamo 2.3 Patch 1 User Manual

Page 1396

Advertising
background image

6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1380

■■■■■■■■

tiamo 2.3 Patch 1

Cylinder serial number

Serial number of cylinder. It is printed on new cylinders. With intelligent
exchange units it is read off automatically. It can be modified at any time,
e.g. if the cylinder has to be replaced.

IEU
Input

8 characters

Default value

'empty'

EU
Input

10 characters

Default value

'empty'

Parameters for preparation
Configures the parameters to be used in the command PREP.

Volume

Volume to be dosed during preparation.

Range

0.00000 ... 99999.99999 mL

Selection

Cylinder volume

Default value

Cylinder volume

Cylinder volume
The whole contents of the cylinder will be dosed.

Cycles

Number of rinsing cycles during preparation.

Range

1 ... 9

Default value

2

Dosing rate

Rate at which the solution is to be dosed. The maximum dosing rate
depends on the cylinder volume of the exchange unit used (see below).
When the function is carried out the rate will automatically be decreased
to the highest possible value.

Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum

Filling rate

Rate at which the cylinder is to be filled. The maximum filling rate depends
on the cylinder volume of the exchange unit used (see below). When the
function is carried out the rate will automatically be decreased to the
highest possible value.

Advertising