5 Method

tiamo 2.3 Patch 1

■■■■■■■■

1133

Temperature
The section Temperature is only displayed for Device type = 089 Pho-
tometer
.

Temperature

Measuring temperature which can be entered manually. If a temperature
sensor is connected and the Temperature measurement is set on the
General/Hardware tab under Sensor to continuous, then the temper-
ature will be measured continuously.

Range

20.0 ... 60 °C

Default value

25.0 °C

5.6.5.12

CAL LOOP pH

5.6.5.12.1

CAL LOOP pH - Overview

Dialog window: Method

CAL LOOP pH Properties... CAL LOOP pH - 'Com-

mand name'

Loop command for the calibration of pH electrodes which is used
together with the command CAL MEAS pH .

Appearance
The command has the following appearance:

Parameters
The parameters for the command CAL LOOP pH are configured in the
following dialog window:

CAL LOOP pH - Properties

Command variables
The following command variables are generated in the method run by the
command CAL LOOP pH and can be used in formulas under the desig-
nation 'Command name.Variable designation':

Desig-
nation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command has
never been started

.ENP

Electrode zero point calculated from the calibration