5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

548

■■■■■■■■

tiamo 2.3 Patch 1

Min. waiting time

The minimum waiting period is only important for drift-controlled meas-
urements. The measured value is not accepted until the minimum waiting
time has elapsed, even if the signal drift has already been reached. The
drift continues to be checked while the waiting time is elapsing.

Range

0 ... 999999 s

Default value

0 s

Max. waiting time

If signal drift has been switched off or has not yet been reached, then the
measured value will be accepted when the maximum waiting time has
elapsed.

Range

0 ... 999999 s

Default value

1 s

Start volume

Start volume

Volume to be added before the start of the titration at the dosing rate
indicated.

Titrando, 855
Range

0.00000 ... 9999.99 mL

Default value

0.00000 mL

Titrino
Range

0.00 ... 999.99 mL

Default value

0.00 mL

Dosing rate

Speed at which the start volume is to be added. The maximum dosing rate
depends on the cylinder volume of the exchange unit or dosing unit used.

Titrando, 855
Range

0.01 ... 166.00 mL/min

Selection

maximum

Default value

maximum

Titrino
Range

0.01 ... 150.00 mL/min

Selection

maximum

Default value

maximum