PDQuest User Guide

L-8

You can draw more boxes than the number of segments defined in the Calstrip
Record. However, they are automatically and permanently deactivated.

L.1.d Plot the Calibration Curve

Once you have created the sampling boxes, select Plot Calib. Curve to display the
calibration curve.

A graph will be displayed that correlates the optical density (usually between 0.00
and 2.00 O.D.) with DPMs or other specified units.

The

Y

axis represents the optical density (0.00–2.00) in each sampled segment, while

the

X

axis represents the corresponding DPM or other unit per image unit area

calculated from the calibration record data.

Fig. L-4. Example of a calibration curve.

Advertising