background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

993

5.6.4.7.4

MEAS T - Evaluations

Tab: Method

MEAS T Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fix measured quantity (Meas-
ured value
or Time) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.7.6, page 997).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.7.6, page 997).

[Delete]

Delete the selected line.

Minimum evaluation

Minimum evaluation

on | off (Default value: off)

If this option is activated, then the associated time for the minimum meas-
ured value will be interpolated from the measuring point list.

Threshold value

The evaluation of the minimum begins as soon as the slope of the curve
exceeds the set threshold value.