7 met cond - additional evaluations, Additional evaluations – Metrohm tiamo 2.3 (process analysis) User Manual

Page 701

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

685

Evaluation with volume window (mL)
With this option, an area (window) can be defined on the volume axis.
The equivalence point is recognized only if it lies within this window
and in addition fulfills the parameters defined in the window.

Evaluation without window

Smoothing

Measure for the curve smoothing used for the data processing. The higher
the smoothing that is selected, the more the measurement curve will be
smoothed.

Range

0 ... 150

Default value

0

Evaluation with volume window (mL)
With this option up to 9 regions (windows) can be defined on the volume
axis. Only those equivalence points which lie within these windows and
additionally meet the parameters defined for each window will be recog-
nized. Only one EP will be recognized per window. The defined windows
with their parameters are shown in the window table and can be edited
with the following buttons:

Table of the volume windows

The parameters for the defined volume window are displayed in the win-
dow table, which cannot be edited. The following columns appear Lower
limit mL
, Upper limit mL and Smoothing.

[New]

Open the dialog window Volume window # in which the parameters
for a new window can be entered.

[Properties]

Open the dialog window Volume window # in which the parameters
for the selected window can be edited (see Chapter 5.6.3.3.6.9, page
690)
.

[Delete]

Delete the window selected in the table.

5.6.3.3.6.7

MET Cond - Additional evaluations

Tab: Method

MET Cond Properties… Additional evaluations

Command name

Name of the command.

Input

25 characters

Advertising