5 dos u - monitoring, Monitoring – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1204

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1188

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Titrino
Range

0.01 ... 150.00 mL/min

Default value

maximum mL/min

Note

The filling rate should be reduced for viscous liquids.

5.6.6.3.5.5

DOS U - Monitoring

Tab: Methods

DOS U Properties... Monitoring

Command name

Name of the command.

Input

25 characters

The following monitoring processes can be switched on and defined on
this tab:

Monitoring Measured value

Monitoring Measured value

on | off (Default value: off)

If this option is switched on then the measured value will be monitored
and any exceeded limits will be entered in the measuring point list.

Lower limit

Lower limit of the measured value. When the measured value falls below
this value, the event Measured value lower limit exceeded is trig-
gered.

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

-2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

-2000 mV

Lower hysteresis

Lower hysteresis of the measured value. If the measured value exceeds the
lower limit by this hysteresis value, then the event Measured value
lower limit ok
is triggered.