8 kfc - additional evaluations, Additional evaluations – Metrohm tiamo 2.3 (process analysis) User Manual

Page 854

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

838

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Automatic start after sample addition

on | off (Default value: off)

If this option is enabled, the titration starts automatically when the com-
mand is in status COND OK and the Threshold value set is exceeded. In
this case the system assumes that the sample has been injected. This para-
meter is displayed only for 851 and 852.

Threshold value

Value the automatic start after the sample addition is carried out automat-
ically when it is exceeded.

851, 852
Range

0 ... 999 mV

Default value

50 mV

Only start titration by a start command from a SEND command

on | off (Default value: off)

If this option is activated, then the titration will not be started until the
Event message Start titration is sent to the command with a SEND
command.

Note

Conditioning must already be completed (= Condok) at the time the
SEND command is received, otherwise the command to start the titra-
tion will be ignored.

If this command is already started (e.g. via Automatic start after
sample addition
), the SEND command will not have any effect.

Show measured value during conditioning

on | off (Default value: off)

If this option is enabled, then the measured value will be shown in the
subwindow Live display of the program part Workplace, in addition to
the drift and the time.

5.6.3.6.8

KFC - Additional evaluations

Tab: Method

KFC Properties... Additional evaluations

Command name

Name of the command.

Input

25 characters