5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1227

Range

1 ... 999 min

Selection

off

Default value

off

Max. waiting time

If the Target temperature is not reached within the defined Heating
time
then this waiting time will start. This parameter is only visible if a
Heating time has been defined.

This value is live editable for the 874.

Range

1 ... 999 min

Selection

off

Default value

off

Action if timeout expires

on | off (Default value: on)

If this option is activated, then an action can be defined which is to be
carried out if the target temperature has not yet been reached after expiry
of the maximum waiting time (is displayed only when Timeoutoff).

Selection

Display message | Cancel determination | Can-
cel determination and series

Default value

Display message

Display message
A message is displayed and you can select whether you want to con-
tinue the run or cancel it.
Cancel determination
The determination is canceled and the next determination of the series
is started.
Cancel determination and series
The determination as well as the series will be canceled, then the exit
track (if present) will be started

Heater (Photometer)

Target temperature

Temperature to which the cuvette has to be heated. With Init.temp. the
cuvette heating is switched on at the same time as the device and the
cuvette is heated up to this temperature.

This value is live editable.

The initial temperature is defined in the device configuration.

Range

20 ... 80 °C

Selection

Init.temp. | off

Default value

Init.temp.