Metrohm tiamo 2.3 (process analysis) User Manual
Page 1379
■■■■■■■■■■■■■■■■■■■■■■
6 Configuration
tiamo 2.3 (for Process Analysis)
■■■■■■■■
1363
Electrode type "Standard"
Parameters
Excellent
electrode
Good
electrode
Usable
electrode
Max. streaming potential
(mV)
2.5
3.0
4.0
Max. drift (mV/min)
2.0
2.5
3.0
Min. slope (%)
96.5
96.0
95.0
Max. slope (%)
101.0
102.0
103.0
Max. response time (s)
45
50
60
Min. offset potential (mV)
-15
-15
-15
Max. offset potential (mV)
15
15
15
Electrode type "Gel"
Parameters
Excellent
electrode
Good
electrode
Usable
electrode
Max. streaming potential
(mV)
3.0
3.5
4.5
Max. drift (mV/min)
2.5
3.0
4.0
Min. slope (%)
96.5
96.0
95.0
Max. slope (%)
101.0
102.0
103.0
Max. response time (s)
60
75
90
Min. offset potential (mV)
-20
-20
-20
Max. offset potential (mV)
20
20
20
Electrode type "Non-aqueous"
Parameters
Excellent
electrode
Good
electrode
Usable
electrode
Max. streaming potential
(mV)
3.0
4.5
6.0
Max. drift (mV/min)
5.0
7.0
9.0
Min. slope (%)
88.0
80.0
70.0
Max. slope (%)
120.0
130.0
140.0
Max. response time (s)
60
75
90
Min. offset potential (mV)
-50
-50
-50