5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1174

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Upper hysteresis pH

The upper hysteresis of the measured value. If the measured value once
again falls below the upper limit by this hysteresis value, then the event
Measured value upper limit ok will be triggered.

Titrando, 855
Range

0 ... 20.000

Default value

20

Action

Selection of the event to be triggered if the upper or lower limit is excee-
ded:

Selection

Stop determination | Stop command (only
Titrando)
| Wait for [Continue] | Wait for limit
ok
| none

Default value

none

Stop determination
The running DOS command is quit, then the exit track (if present) is
started and the determination is stopped.
Stop command (only Titrando)
The running DOS command is quit, then the next command is execu-
ted.
Wait for [Continue]
Reagent dosing in the current DOS command is interrupted and a
message appears. As soon as the monitored measured value is again
within the limits (including hysteresis) reagent dosing can be resumed
by pressing the [Continue] button in this message box.
Wait for limit ok
Reagent dosing in the current DOS command is interrupted. As soon
as the monitored measured value is again within the limits (including
hysteresis), reagent dosing will be resumed automatically.
none
No action will be taken if limits are exceeded.

Monitoring Temperature

Monitoring Temperature

on | off (Default value: off)

If this option is switched on then the temperature will be monitored and
any exceeded limits will be entered in the measuring point list.