6 cal loop conc, 1 cal loop conc - overview, Cal loop conc – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1121

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1105

5.6.5.6

CAL LOOP Conc

5.6.5.6.1

CAL LOOP Conc - Overview

Dialog window: Method

CAL LOOP Conc Properties.. CAL LOOP Conc -

'Command name'

Loop command for the calibration of ion-selective electrodes, which
is used together with the command CAL MEAS Conc.

Appearance
The command has the following appearance:

Parameters
The parameters for the command CAL LOOP Conc are configured in the
following dialog window:

CAL LOOP Conc - Properties

Principle
For calibrating ion-selective electrodes (ISE), a linear correlation of the
measured potential U

i

and the logarithm of the sought ion concentration

c

i

cannot always be assumed. In ion measurements, interfering ions have

an influence on the measurement and that is why this effect is perceptible
by a flattening of the curve, especially in low concentration ranges.

The influence of interfering ions can be described by the Nikolski equa-
tion. This is an extended Nernst equation which considers the selectivity of
the ISE. For the ISE calibration with tiamo it is used in the following form: