Metrohm tiamo 2.3 (process analysis) User Manual

Page 599

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

583

EP recognition

Filter for the recognition of equivalence points:

Titrino
Selection

all | greatest | last | off

Default value

all

Titrando, 855
Selection

all | greatest | last | ascending | descending |
off

Default value

all

all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point to be found will be recognized.
ascending
Only equivalence points with a positive slope of the titration curve will
be recognized.
descending
Only equivalence points with a negative slope of the titration curve will
be recognized.
off
Equivalence point recognition is switched off.

Evaluation with measured value window (pH)

[New]

Open the dialog window Measured value window # in which the
parameters for a new window can be entered (see Chapter 5.6.3.3.2.9,
page 589)
.

[Properties]

Open the dialog window Measured value window # in which the
parameters for the selected window can be edited (see Chapter
5.6.3.3.2.9, page 589)
.

[Delete]

Delete the window selected in the table.

Advertising