6 Configuration

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1435

Warning limits for E(0) (for ISE electrodes only)
on | off (Default value: off)

If this option is activated, then the values for E(0) will be shown in
orange if the limits are exceeded.

Lower limit

Lower warning limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

-15 mV

Upper limit

Upper warning limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

15 mV

Intervention limits for E(0) (for ISE electrodes only)
on | off (Default value: off)

If this option is activated, then the values for E(0) will be shown in red if
the limits are exceeded.

Lower limit

Lower intervention limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

-30 mV

Upper limit

Upper intervention limit for E(0).

Range

-2.147E+9 ... 2.147E+9 mV

Default value

30 mV

Warning limits for cell constant (for conductivity sensors
only)
on | off (Default value: off)

If this option is activated, then the values for the cell constant will be
shown in orange if the limits are exceeded.

Lower limit

Lower warning limit for the cell constant.

Range

0.001 ... 500 /cm

Default value

2.0 /cm