5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1102

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Calibration wavelengths
Input of the reference wavelengths. Definition of the behavior of the
calibration.

Command variables
The following command variables are generated in the method run by the
command CAL Spec and can be used in formulas under the designation
'Command name.Variable designation':

Designation

Description

.BSY

Command status; 1 = BUSY, HOLD or ERROR; 0 =
READY
; invalid (variable not available) = Command
has never been started

.C0

Coefficient c0 of the calibration curve

.C1

Coefficient c1 of the calibration curve

.C2

Coefficient c2 of the calibration curve

.C3

Coefficient c3 of the calibration curve

.CAL

Status of the calibration: 1 = calibrated normally, 0
= canceled
, because reference wavelength was not
found.

.DBL

Total duration for the processing of the command in s

.STY

Type of stop with which the command was stopped: 1
= stopped normally
, 0 = stopped manually with
[Quit]
or [Stop] .

.CP{#}.MEA

Measured wavelength of the calibration point in nm. If
no measured peak can be assigned, then not found
will be entered.

.CP{#}.REF

Reference wavelength of the calibration point in nm.

.FIN

Command status: 1 = Command has been ended
at least once
, 0 = Command was started but not
ended
, invalid = Command was not started (vari-
able not available)
.

.TOL

Tolerance within which a measured peak can be
assigned to a calibration wavelength.