5 titer history limit values – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1414

Advertising
background image

6.6 Subwindow Titrants/Solutions

■■■■■■■■■■■■■■■■■■■■■■

1398

■■■■■■■■

tiamo 2.3 (for Process Analysis)

History graph
The graph shows the last 10 titer values for the selected solution. The titer
values are shown in different colors in the history table:

Blue, if the titer has been determined automatically by a method.

Black, if the titer has been entered manually.

If limits have been defined, then the warning limits will be shown in
orange and the intervention limits in red.

Note

The history graph can be copied into the clipboard by using the con-
text-sensitive menu item Copy.

[Limits]

Open the dialog window Limits for titer, in which the warning and inter-
vention limits can be defined for the titer (see Chapter 6.6.3.5, page
1398)
. These limits apply only to the graph, no monitoring is carried out
during the titer determination.

[Delete history]

Delete the titer history.

6.6.3.5

Titer history Limit values

Dialog window: Configuration

Titrants/solutions Properties Titer his-

tory

[Limit values] Limit values titer

Warning and intervention limits can be defined for the titer in the dialog
window Limits for titer. If you have defined limits then these will be
shown in the graph in orange for warning limits and red for intervention
limits. Whether these limits are observed is not monitored, i.e. if the limits
are exceeded no action will be taken and these solutions can be used in
determinations.

Warning limits for titer
on | off (Default value: off)

If this option is activated, then the titer value will be shown in orange if
these limits are exceeded.

Lower limit

Lower warning limit.

Input

10 digits

Default value

0.9750

Advertising