5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

677

Sensor

Selection of a sensor of the Conductivity sensor type from the sensors
configured in the sensor table. The calibration data for the sensor will be
adopted for the determination.

Selection

Sensor name | Conductivity sensor

Default value

Conductivity sensor

Temperature measurement

Type of temperature measurement.

Selection

continuous | automatic | off

Default value

automatic

continuous
A temperature sensor must be connected. The temperature will then
be measured continuously.
automatic
If a temperature sensor is connected then the temperature will be
measured continuously. Otherwise the temperature entered manually
under Temperature on the Titration parameters tab will be used.
off
The temperature will not be measured. The temperature entered man-
ually under Temperature on the Titration parameters tab will be
used.

Stirrer

Stirrer

Selection of the stirrer.

Selection

1 | 2 | 3 | 4 | off

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. The algebraic sign of the stirring rate changes the
direction in which the stirring is done.

Range

-15 ... 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.