Metrohm tiamo 2.3 (process analysis) User Manual

Page 1132

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1116

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Buffers

Number of buffers

Displaying the number of buffers (3), with which the electrode test is to be
carried out.

Buffer type

Selecting a predefined buffer series for the pH calibration. It is imperative
that the selected buffer series include the buffers 4, 7 and 9.

If these buffers are used, tiamo will recognize the particular buffer auto-
matically.

Selection

Metrohm | NIST | DIN | Fluka Basel | Mettler |
Merck CertiPUR 20°C | Merck CertiPUR 25°C |
Radiometer | Baker | Hamilton DURACAL | Pre-
cisa

Default value

Metrohm

Request for buffer exchange

on | off (Default value: on)

If this check box is activated, a request for manual buffer exchange is dis-
played after each time the calibration loop is run through. This parameter
must be disabled for automatic buffer exchange when calibrating with a
Sample Processor.

Electrode test

Electrode type

Selection of a predefined electrode type which is to be used to assess the
pH electrode.

Selection

Standard | Gel | Non-aqueous | 'Electrode type
name'

Default value

Standard

'Electrode type name'
Uses the own electrode type defined under this name in the configura-
tion under Extras

Templates Electrode type.

Save calibration data

on | off (Default value: off)

If this check box is activated, then the Slope and pH(0) calibration data
determined at the time of the electrode test is stored in the configuration
with the sensor used.