8.4 Manual control - Functions

■■■■■■■■■■■■■■■■■■■■■■

1656

■■■■■■■■

tiamo 2.3 (for Process Analysis)

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Filling rate

The buret will be filled at this rate after the dosing. The maximum filling
rate depends on the cylinder volume of the exchange unit or dosing
device used. If the entered filling rate is too high for the selected dosing
device, it will automatically be reduced during dosing to the largest possi-
ble value.

Note

The filling rate should be reduced for viscous liquids.

Selection

maximum

Default value

maximum

Titrando, 814, 815, 855
Range

0.01 ... 166.00 mL/min

730, 774, 778, 789
Range

0.01 ... 160.00 mL/min

Titrino
Range

0.01 ... 150.00 mL/min

Fill automatically

on | off (Default value: on)

This parameter is only visible for devices of the type Titrando, Dosing Inter-
face and USB Sample Processor.

If this option is activated, the buret is filled automatically after dosing.
During the filling procedure the volume display is reset to 0.0000 mL. If
this option is deactivated, the added volume is displayed cumulatively.

[Start]

Start fixed volume dosing for the selected dosing device. The dosed vol-
ume is displayed live.