5 properties - data import, Data import – Metrohm tiamo 2.3 (process analysis) User Manual

Page 159

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

3 Workplace

tiamo 2.3 (for Process Analysis)

■■■■■■■■

143

3.6.3.9.5

Properties - Data import

Tab: Workplace

Run Single determination Properties Run window Prop-

erties - Single determination

Data import

Enabling/disabling the import of external data for fields on tab Single
determination
.

Note

The successful conclusion of the data import during the run will be sig-
naled by an acoustic signal (beep).

Import from balance
on | off (Default value: off)

If this option is enabled, then external data from the selected balance will
be imported into the fields on the tab Single determination. Data can
be imported into the fields ID1…ID4, Sample size and Unit.

Note

In order to be able to import sample identifications from the balance,
data export must be set accordingly at the balance (Use designations
ID1, ID2, ID3, ID4).

Balance

Selection of the balance from which data is to be imported.

Selection

Device name

Selection

Always accept data | Accept data only for
active window

Default value

Always accept data

Always accept data
If this option is selected, then the data sent from the balance is impor-
ted into all of the Single determination tabs of the open workplaces
for which this balance is defined as data source.
Accept data only for active window
If this option is selected, then the data sent from the balance is impor-
ted only into the Single determination tab of the workplace that is
actively selected.