3 meas opt conc- measuring parameters – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1070

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

1054

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Switch off automatically

on | off (Default value: on)

If this check box is activated, the stirrer will be switched off automatically
when the command has finished.

5.6.4.15.3

MEAS Opt Conc- Measuring parameters

Tab: Method

MEAS Opt Conc Properties... Measuring parameters

Command name

Name of the command.

Input

25 characters

Tab for the Avantes spectrometer
The parameters for the measuring procedure can be defined on this tab.

Wavelength

Wavelength at which the measured quantity (absorbance) is recorded

Range

100.0 ... 2000.0 nm

Default value

400.0 nm

Blank value

Value by which the measured quantity (absorbance) is corrected. Here an
expression can also be entered through the formula editor (e.g. xx.EME
from other measuring command) by which every measured absorbance
value can then be corrected.

Range

-1E+12 ... 1E+12 (max. 10 digits)

Default value

1.000

Unit

Concentration unit for the blank value.

Input

1 ... 10 characters

Selection

mol/L | mmol/L | µmol/L | g/L | mg/L | µg/L |
mg/mL | µg/mL | ppm | % | mEq/L

Default value

ppm

Integration time

Integration time for adsorption measurement.

Range

0.01 ... 600000 ms

Default value

6 ms

Advertising