4 meas u - evaluations, Evaluations – Metrohm tiamo 2.3 (process analysis) User Manual

Page 970

Advertising
background image

5.6 Commands

■■■■■■■■■■■■■■■■■■■■■■

954

■■■■■■■■

tiamo 2.3 (for Process Analysis)

General/Hardware tab under Sensor to automatic or continuous,
then the temperature will be measured continuously.

Titrando, 855, 867
Range

-20.0 ... 150 °C

Default value

25.0 °C

Titrino
Range

-170.0 ... 500 °C

Default value

25.0 °C

5.6.4.4.4

MEAS U - Evaluations

Tab: Method

MEAS U Properties... Evaluations

Command name

Name of the command.

Input

25 characters

The following additional methods for evaluation of measurement curves
can be activated and defined on this tab.

Fixed endpoint evaluation

Fix endpoint evaluation

on | off (Default value: off)

If this option is activated, then the associated values for the other meas-
ured quantities will be interpolated with a fix measured quantity (Meas-
ured value
or Time) for the fix endpoint from the measuring point list.

Moves the selected line up (changes the sequence).

Moves the selected line down (changes the sequence).

[New]

Open the dialog window Fixed endpoint evaluation # to enter a new
fixed endpoint (see Chapter 5.6.4.4.6, page 958).

[Properties]

Open the dialog window Fixed endpoint evaluation # to edit the fixed
endpoint selected in the table (see Chapter 5.6.4.4.6, page 958).

[Delete]

Delete the selected line.

Advertising