7.16 089 Photometer

■■■■■■■■■■■■■■■■■■■■■■

1584

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Stirrer 1

Stirrer type

Display of the stirrer type.

Stirrer serial number

Display of the serial number of the connected stirrer.

Remote box 1

Display of the connected remote box.

7.16.6

Properties - GLP

Tab: Configuration

Devices [Edit] Properties... Properties - 'Device type' -

'Device name'

GLP

GLP test date

Date of the last GLP test. This date can be selected by clicking on

in

the dialog window Select date (see Chapter 2.5.1, page 89).

Comment on GLP test

Comment on GLP test.

Input

1000 characters

Monitoring of GLP validity

Monitoring of GLP validity

on | off (Default value: off)

If this option is enabled then the time interval for the GLP test will be
monitored.

Monitoring can only be switched on when a date has been entered in
the field GLP test date.

GLP test interval

Time interval to next GLP test. If a value is entered here then the date in
the field Next GLP test will be adjusted automatically.

Can only be edited if monitoring is switched on.
Range

1 ... 999 days

Default value

999 days

Next GLP test

Date on which the next GLP test is to be carried out. The date can be

selected by pressing

in the dialog window Select date (see Chapter