5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

535

Titrando, 855
Range

-2000.0 ... 2000.0 mV

Default value

2000.0 mV

Titrino
Range

-2000 ... 2000 mV

Default value

2000 mV

EP criterion

Criterion for the recognition of endpoints. Endpoints whose discovered
ERC value is smaller than the value entered here will not be recognized.

Titrando, 855
Range

0 ... 200

Default value

5

EP recognition

Filter for the recognition of equivalence points:

Titrando, 855
Selection

first | greatest | last | ascending | descending

Default value

first

first
Only the first endpoint to be found will be recognized.
greatest
Only the endpoint with the greatest ERC value, i.e. the steepest jump,
will be recognized.
last
Only the last endpoint to be found will be recognized.
ascending
Only endpoints with a positive slope of the titration curve will be rec-
ognized.
descending
Only endpoints with a negative slope of the titration curve will be rec-
ognized.

5.6.3.2.4.10

DET Ipol - Volume window

Dialog window: Method

DET Ipol Properties... Potentiometric evalua-

tion

Evaluation with volume window (ml) [New]/[Properties] Volume win-

dow #

Volume windows are defined sub-regions (windows) on the volume axis
for which different parameters for the potentiometric evaluation can be
defined. Only those endpoints which lie within these windows and addi-