5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

577

Titration rate

Titration rate

The three predefined sets of parameters can be selected for the titration
rate: slow, optimal and fast; the parameters for the Measured value
acceptance
and Dosing of increments of these sets are not displayed.
The User setting must be selected in order to be able to edit these param-
eters.

Selection

slow | optimal | fast | User

Default value

optimal

slow
For titrations in which the finest details are also to be visible. This could
however also lead to an increase in noise, which could result in unwan-
ted EPs.
optimal
Parameter set for all standard titrations; optimized for the most fre-
quent applications.
fast
For less critical rapid titrations.
User
Editing the individual titration parameters which affect the titration
rate.

Measured value acceptance

Note

Will be displayed only if the option User is selected in the selection list
Titration rate.

Measured value drift

Drift for the measured value acceptance during the titration.

Titrando, 855
Range

0.1 ... 999.0 mV/min

Default value

20.0 (slow), 50.0 (optimal), 80.0 (fast) mV/
min

Selection

off

off
Measured value acceptance will take place after the maximum waiting
time has elapsed.