4 properties - titer history, Titer history – Metrohm tiamo 2.3 (process analysis) User Manual

Page 1413

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

6 Configuration

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1397

6.6.3.4

Properties - Titer history

Tab: Configuration

Titrants/solutions Properties Titer history

History table
The table contains the last 10 titer determinations for the selected solution
and can neither be edited nor sorted. The titer determinations are
arranged chronologically so that the most recent determination is shown
last.

Titer

Titer values are shown in the following colors:

Blue, if the titer has been determined automatically by a method.

Black, if the titer has been entered manually.

Orange, if the titer is outside the warning limits.

Red, if the titer is outside the intervention limits.

If the warning or intervention limits are exceeded then the line number
will also be shown with an orange and red background respectively.

Date titer det.

Date and time of the titer determination.

Titer method

Name of the method with which the titer determination was carried out.
If the titer has been entered manually then manual will be shown here.

Note

This parameter is not saved on the data chip of intelligent exchange/
dosing units, i.e. this parameter will be empty if the data are written
into the solutions table from the IEU/IDU.

User

Short name of the user logged in during titer determination or when the
titer was entered manually. If work is not carried out using login then the
user logged in under Windows will be entered automatically.

Note

This parameter is not saved on the data chip of intelligent exchange/
dosing units, i.e. this parameter will be empty if the data are written
into the solutions table from the IEU/IDU.

Advertising