5 Method

tiamo 2.3 (for Process Analysis)

■■■■■■■■

1091

* Definition of the number of calibration solutions and the type of solution
exchange.

** Move to the next calibration solution.

 
The request for changing the buffer or standard must be deactivated in
the command CAL LOOP pH or CAL LOOP Conc, respectively. No
request for changing the buffer or standard is then displayed during the
run.

For the automatic move to the next calibration solution on the Sample
Processor, a MOVE command must be inserted between the loop com-
mand CAL LOOP and the measuring command CAL MEAS. The posi-
tions of the calibration solutions on the rack are defined in this command.
The following three possibilities are available to do so: