6.3 Configuration data

■■■■■■■■■■■■■■■■■■■■■■

1350

■■■■■■■■

tiamo 2.3 (for Process Analysis)

Backup name

Selection of a backup file.

Selection

'Backup files'

Backup date

Display of the moment at which the configuration database was backed
up. If the backup file is located on a network drive, this information is not
available.

Database name

Display of the name of the configuration database. If the backup file is
located on a network drive, this information is not available.

Size

Display of the size of the configuration database in KB.

Save as

Display of the name under which the configuration database will be
restored.

[Start]

Start the recovery of the configuration database. After the start, a pro-
gress bar appears in the window. When the backup has been completed,
the dialog window closes automatically.

6.3.3

Templates

6.3.3.1

Templates - Custom calibration buffers

6.3.3.1.1

Managing templates for custom calibration buffers

Dialog window: Configuration

Tools Templates Custom calibration buffer

You can define 9 of your own calibration buffers for pH calibration with
automatic buffer recognition in the dialog window Custom calibration
buffers
. This buffer series is globally valid and can be selected as buffer
type Custom in the command CAL LOOP pH.

Buffer table
The buffer table shows the defined pH values for the calibration buffers in
5 °C increments. The table can neither be edited nor sorted.

[Edit]

The dialog window Change pH value opens.

[Delete all]

All the pH values of the selected buffer will be set to off.

Advertising